3.2 The Calibration Utility
3.2.1 General
The calibration utility (LCIC-WIM-CALIBRATION) enables to calibrate
the LCIC-WIM board adjusting it to your own system.
The utility is straightforward and is in the form of a Windows wizard. It
includes three main stages carried out in five steps.
The three stages are:
1. Show Data (step 1)
(Described below under ‘Step 1’.)
2. Pseudo Calibration (steps 2-3-4)
The calibration is called ‘pseudo’ as it won’t be saved in the
board unless the user confirms it in the next stage. As long
as the user did not confirm the new (pseudo) calibration, the
previous calibration remains in effect.
(The details are described below under ‘Step 2-3-4’.)
3. Save or Quit (step 5)
(Described below under ‘Step 5’.)