5
Models: HX93AV, HX93AV-D HX93AV-RP1 (Voltage Output)
TB1 - Relative Humidity TB2 - Temperature
1. + Power Supply 1. + Power Supply
2. + Output 2. + Output
3. - Power Supply 3. - Power Supply
7. Voltage Transmitter Wiring Examples
For voltage output models (0 to 1 Vdc)
Figure 5 - Voltage Transmitter Wiring Example
8. Relative Humidity Output Calculations
To calculate % Relative Humidity by measuring the current or voltage
output use the following formulas.
For current output:
% RH = (Current measured in miliamps – 4) ÷ .16
EXAMPLE: (11.04 mA – 4) ÷ .16 = 44% RH
For voltage output:
% RH = (Voltage measured in volts x 100)
EXAMPLE: .44 x 100 = 44% RH
Comentarios a estos manuales